Based on the photovoltaic industry, with optical, mechanical, electronic based on the combination of the three core technologies

Gas Chromatography

 >Products>Analytical Instruments>Gas Chromatography

  • Product introduction
  • Technical specifications
  • Download
  • Sales contact

Flexible Sample Introduction System

Up to three injectors may be installed and operated simultaneously with independent temperature control. Wide range of injectors is available to be chosen from for various applications. It is very convenient for installation and disassembly of sample injectors.
Sample Injector with Mono Packed Column/Double Packed Columns:
Several alternative sample introduction methods, including on-column sample introduction and six-way valve injection;
Adapter for injection to capillary column (0.53 mm i.d.).
Sample injector for capillary column
Special sample injector with septum purge and back-pressure valve splitting;
Special programmable split/splitless injector with solenoid valve, septum purge, back-pressure valve splitting, and standalone carrier gas and make-up gas flow;
Modified dual-packed-column sample injector for injection to capillary column by employing some connectors and back-pressure valve splitting.


Advanced Pneumatics
 

Carrier gas control with constant pressure valve and constant flow valve;


Air and hydrogen control with constant pressure valve and needle valve;


Flow easily adjustable through constant flow valve and needle valve with digital display;


High precision, reproducibility and reliability.


Sensitive and reliable detector
 

Up to three detectors (FID, TCD, ECD, etc) may be installed and operated simultaneously. Simultaneous multi-detection with detectors in series or in parallel meets the various analysis requirements.


The new FID provides higher sensitivity than ever before. It is very convenient to disassemble and clean the jet and collector. Precisely fixed emitter configuration ensures conformity of product performance;


Excellent TCD detector with high-resistance tungsten-rhenium filaments provides high sensitivity. A sensitivity of 10000 mV•mL/mg is achieved with optional amplifier circuit board. Special configuration of air heat insulation enhances the stability of TCD detector


Constant-current, pulse-modulated ECD detector sets sensitivity and current by computer.


Advanced Microcomputer Control System

With advanced control techniques, microcomputer-based temperature control system provides superior performance.


High temperature precision (± 0.5℃), high reliability, and high resistance to disturbance;


Five independent heated zones, with maximum operating temperature of 400℃;


Temperature-limit setting and overheat protection.


Instrument control is accomplished through the keypad, and the instrument offers outstanding capabilities. Self-diagnosis;


Power-off protection;


Ten methods saving and reloading;


Two controls of external events;


Timing on/off of temperature control;


Timing program, detector status, measurement range, polarity and current setting and display;


Temperature setting values, actual values, retention time and analysis time display.


High-performance spacious column oven

Spacious column oven accommodating capillary column and dual packed columns and convenient for column installation;


Fast heat-up and rapid cool-down column oven with automated back-door opening (300℃ to 50℃ in 7 min);


Available for both near-ambient and above ambient temperature control;


Supports 5 oven ramps with 6 plateaus.

 

Temperature Control

Temperature Range:7℃ ~ 400℃above room temperature, (increment 1℃) High Sensitivity Thermal conductivity detector(HTCD)

Sensitivity: ≥10000mV﹒mL/mg (C16)

Tempperature Control Object:Column Box, front and backdetecor, front and back sampler, assist,total six channels. Electron capture detector(ECD) Detectability: ≤1×10-13g/s  (γ-666)
Program ramps : 7 ramps in total Linearity range: ≥103
Program setting:0.1℃~40℃/min(increment 0.1℃) Maximum using temperature:  350℃
Total time of total program : Up to 655 minutes, (increment 0.1℃) Flame photometric detector(FPD) Detectability: ≤2×10-11g/s(P) 
(Parathion-methyl)≤1×10-10g/s(S)  (Parathion-methyl)

Hydrogen flame ionization detector (FID)

Minimum detectable quantity: M≤8×10-12g/s (C16) Nitrogen-phosphorus detector(NPD)

Detectability: ≤5×10-12g/s(N)  (Azobenzene)≤5×10-13g/s(P)  (Malathion)(On programmed)

Best test result: M≤3×10-12g/s (C16) Photoelectric ion detector(PID) Detectability: ≤1ppb  (Benzene)
Noise: ≤5×10-14A    
Drift: ≤6×10-13A/h    
Linearity range: ≥106
   

Thermal conductivity detector(TCD)

Sensitivity:≥5000mV﹒mL/mg (C16)  

 

Noise: ≤20µV    
Drif: ≤30µV/h    
Linearity rang: ≥104    

 
Optional Parts

Main function

GC1120-TCD

Thermal conductivity detector
GC1120-ECD

Electron capture detector

GC1120-FPD

Flame photometer detector
GC1120-NPD

Electron capture detector

Chromatographic work station

External interface box + work station (not include PC)
Simulation software

Imitation find chromatographic condition, emulationteaching, train and knowledge base.

Sales Contact informations:

SHANGHAI SUNNY HENGPING INSTRUMENT CO.,LTD
ADD:5-6/F,Building 8,456 Hong Cao Road,Xu Hui District,Shanghai
P.C:200233
TEL:0086-021-64951010
E-mail:sales@hengping.com
 

IT consultation Business

Network

Sales Network

more

User Login

UserName:
Password:
Forget Password?
If you have not registered, please Register

UserLogin

UserName:
Email:
Phone:
Role:Non Candidates
Password:
Confirm Password:
If you are already a member, please

Forget Password

Confirm Email:


to:input you register email

If you have not registered, please Register

Forget Password

New Password:
Confirm Password:
Code:
If you have not registered, please Register

Please confirm whether the mailbox right

dderet@qq.com

The mailbox is not correct?

Name:
Email:
The Phone:
Note: